- 30 -
(3) Setting the barometric pressure
The barometric pressure only needs to be set once, not for each test
(
9
)
Note (
9
)
Periodic corrections with a barometer are necessary.
(4) Setting Temp CUT and Volume CUT
Select a mode number on the STANDBY screen shown in Fig. 5 on Page 18 to set or
change the mode parameters, and press the FUNCTION key
(
10
).
The screen shown in
Fig. 12 is displayed.
Note (
10
)
The Test Parameters screen cannot be displayed during testing.
<Test Condition> MODE(01) GASOLIN(300)
Sample OFF G-000001 Const A 463
Operator OFF
TANAKA Const B 27
Loss Corr GRAPH
Temp Unit 0.1
o
C (
o
C) -> (%) Print
Vol Unit
0.1 %
1 100.0 5 200.0
Manu.Heat
OFF
2 - - - - 6 - - - -
Both EP&DP OFF
3 - - - - 7 - - - -
Hesi.Corr
OFF 4 - - - -
8 - - - -
Prnt.Result
OFF (%) -> (
o
C) Print
Temp Cut
- - - -
o
C
1 50.0 5 100.0
Vol Cut - - - -
%
2 - - - - 6 - - - -
IT&Rate Auto Set OFF
3 - - - - 7 - - - -
Baro Method D850 4 - - - - 8 - - - -
Const A does not appear and Const
B will be changed to Const K when
Baro Method is D1078.
Both Const A and Const B don’t
appear when Baro Method is D86.
Fig.12 Test Parameters Screen
(a) Temp Cut, Vol Cut, and (
°
C)
→
(%) Print and (%)
→
(
°
C) Print display
▪
▪
▪
▪
▪
instead of
000.0 to indicate that no value is set. To set data, position the cursor on one of the
▪
▪
▪
▪
▪
and press the
↑
or
↓
keys. If numeric data is displayed, change as necessary.
(b) After setting data, press the FUNCTION key again to return to the STANDBY screen.
(5) Setting sample or operator names
(a) When setting a sample or operator name, display the Test Parameters screen shown in
Fig. 13 in the same way as (4).
(b) To include the sample or operator name in the printout, change from OFF to ON (refer to
the example of the test results printout shown in Fig. 18 on Page 44).
(c) The following characters can be entered in the Sample or Operator fields:
Blank ! “ # $ % & ‘ ( ) * + ,
- . / 0 1 2 3 4 5 6 7 8 9
A B
C D E F G H I J K L M
N O
P Q R S T U V W X Y Z
(d) If only the
↑
or
↓
keys are used to enter characters in the Sample or Operator fields, it
takes a lot of time to enter. To save some time, the tester provides the following function:
Each time the MANUALCUT key is pressed, a blank, 0, A or N are displayed, in this
order. Select the one closest to the necessary character and then select the intended
character using the
↑
or
↓
key.
(e) After setting the name, press the FUNCTION key again to return the display to the
STANDBY screen.
(6) Setting a mode name
(a) A mode name can be set on the STANDBY screen shown in Fig. 5. Select a mode
number using the operations explained in (1) and press either the
←
or
→
key to display
the short cursor.
(b) Enter a mode name in the same way as (c) and (d) of (5) in Section 9.2: How to Set Test
Parameters on Page 29.
(c) After setting a mode name, press the ENTER key to change to the long the cursor.
Summary of Contents for AD-6
Page 2: ...1 Blank Page...
Page 9: ...8 Blank Page...