Calibration
------------------------------------------------------------------------------------
185
Offset/Full-Scale Calibration of Output Current Monitor(ZX Series Type-LAN)
1.Perform load with only the shunt resistor.
Disconnect the load wires and the remote sensing wires
.
Connect the shunt to the output terminal. Connect the
digital multimeter to the voltage terminal located on the
shunt.
Connect the digital multimeter to the external control
connectors number 3 and 4. (Current Monitor Terminal)
Turn the POWER switch ON and OUTPUT OFF in
advance.
2 . Press the FUNC. key and change the top Number
Display to Settings Item 13 with the VOLTAGE dial.
3.Turn the CURRENT dial and set the current value to 0
(Linearity Compensation: OFF).
4.Press the FUNC. key again to finalize settings.
5.Press the PRESET key ① and set to “1.00V” with
the VOLTAGE dial ② and “40.00A~160.0A” (※)
with the CURRENT dial ③.
※For 400LN/LAN set to “40.00A”
For 800LN/LAN set to “80.00A”
For 1600LN/LAN set to “160.0A”
6.Press the PRESET key ①
7. Set the load switch S1 to OPEN.
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Note:
Current is 40A for
(400LN/LAN)
800LN/LAN is 80A
1600LN/LAN is 160A
Use wires which will not
produce a voltage drop
below 0.8V.
+
-
Digital
Multimeter No.1
Power
Supply
V
Shunt
-
+
Calibration Circuit
V
+
-
3
4
Digital
Multimeter No.2
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
DISPL AY
A
B
C
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
①
②
③
ZX-S-400LN
Summary of Contents for ZX-S-1600LAN
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Page 13: ...About this Product 10 This is blank page...
Page 63: ...Useful Function 60 This is blank page...
Page 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Page 157: ...Special Loads 154 This is blank page...
Page 199: ...Specifications 196 This is blank page...
Page 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Page 201: ...Appendix 198 ZX S 800LN 800LAN...
Page 202: ...Appendix 199 ZX S 1600LN 1600LAN...