Using in Series/Parallel Operation
------------------------------------------------------------------------------------
126
Operation
1.Setting Slave Unit
① Press the Function “FUNC.” key.
② Select Item 73 with VOLTAGE dial.
③ Change the set value to 1 (Slave Mode) with CURRENT dial.
④ Press the FUNC. key again to finalize settings.
⑤ Turn the POWER switch OFF.
※Device will operate as Slave unit when the POWER switch is turned ON again.
In so doing, the “Output” lamp will light.
When returning to Master, set FUNCTION Item 73 to 0,
Item 51 to 0, and Item 60 to 1.
MEMO