Calibration
------------------------------------------------------------------------------------
166
8. Press the FUNC. key and change the top Number
Display to Settings Item 28 with the VOLTAGE dial.
9. Adjust so that the digital multimeter (used for
output voltage measurement) displays “0.00V” with
the CURRENT dial.
10. Press the FUNC. key to complete offset
calibration of output voltage control with external
resistance (A).
11. Set external resistance to 0Ω(short).
12.Press the FUNC. key and change the top Number
Display to Settings Item 29 with the VOLTAGE
dial.
13.Adjust so that the digital multimeter (used for
output voltage measurement) displays to “80.00V”
with the CURRENT dial.
14. Press the FUNC. key to complete full-scale
calibration of output voltage control with external
resistance (A).
15. This completes offset/full-scale calibration of
output voltage control with external resistance
(A).
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
IN
OUT
18
6
外部抵抗
0~∞
Ω
External
Resistance
0~∞Ω
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Summary of Contents for ZX-S-1600LAN
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Page 13: ...About this Product 10 This is blank page...
Page 63: ...Useful Function 60 This is blank page...
Page 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Page 157: ...Special Loads 154 This is blank page...
Page 199: ...Specifications 196 This is blank page...
Page 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Page 201: ...Appendix 198 ZX S 800LN 800LAN...
Page 202: ...Appendix 199 ZX S 1600LN 1600LAN...