Instrument Operating Manual
CyFlow® Cube 8 Rev-27_2015-07-02
Sysmex Partec GmbH
Am Flugplatz 13, 02828 Görlitz, Germany · Phone +49 3581 8746-0 · Fax +49 3581 8746-70 · [email protected] ·
www.sysmex-partec.com
18 / 43
4.1.6
MEASURE
– definition of measure modes
Measmode
allows selection of:
Analyze all:
run sample until its levels reached the stop electrode
Volumetric counting (VC) with volume
a volume can be pre-selected
Volumetric counting (VC) with electrodes:
START and STOP electrodes of the sample port are
used to define a fixed sample volume of 200µl
Cells in region
a particle number can be pre-selected
Speed values
and
Volume values
can be edited and will be stored upon saving a
configuration file.
Autoflags
allows activation of specific functions:
Autostart: automatic start of the measurement
Activate the AutoStart flag
Press start (only once to confirm the process)
Connect a sample tube
Autostart:
With each new sample tube the next measurement will be started until the STOP button is
pressed.
AutoPA:
automatic peak analysis, at the end of the measurement
→
only for Ploidy
Autosave:
Automatic storage at the end of the measurement. Autosave does not function when the
measurement is finished by the “end” button.