6.3.5 BitRange
The values are ranging from 6 to 12. It sets the channel resolution from 16bit
(64channels) up to 12bit (4096 channels).
6.3.6 CR-Mode
This option allows selecting the part of the data to be displayed:
All events
Region only
Color gating
6.4
Region/ROI properties
6.4.1
Create a polygonal region
By double left click on the plot where the gate/region is required, a
first point will be set; a second click will set a further point, and so
on. Label as many points as required with the left mouse button.
To close the gate surface, click mouse right.
6.4.2 Region properties
To access the region property,
Alt+right-click
on the gate to edited. On the
PLOTS
register on the right of the screen a selection of region options is displayed.
Region name
As default, the gate will be named: region n
x
,
successive region: region n
x+1
The user can also define a name. Remember to
validate by
Accept
before to move to another
region.
Home Plot name
This defines the physical plot where the region is
located.
Color RGB
Allows to choose the color of the region. This color will be reported to each plot selected
as
color gating only
.
Max count
This option allows fixing the maximum number of events from a particular ROI (Region Of
Interest) to be analysed (see CyView
™ 8 FILE register).
Sorter region
To set the ROI/region as a sorting gate, just by selecting the box.
(This option is only available on the Cube equipped with Sorter flow cuvette).