5.2
Work process
Choose the work batch for measure and acquire of data.
5.3
Shut down process
For the cleaning of the sample port a separate cleaning mode is available. The detail
procedure would be the following:
Select the “CLEAN” batch by clicking on the “CLEAN” batch
symbol
After pressing Start
the system will start the priming process automatically
and guide through the program.
Connect a sample tube with cleaning solution
(green solution, Order No. 04-4009 )
and
press
“Continue”
Connect a sample tube with decontamination solution
(violet solution, Order No. 04-
4010)
and press “Continue”
Connect a sample tube with sheath fluid
(Order No.04-4007)
and press “Continue”