7.
Typical Sample Analysis
A step by step procedure and what are the important steps in the data acquisition.
Starting a measurement
Load a sample file. The respective instrument settings are automatically loaded.
Or
Load a CFG-file, saved instrument settings will be load.
Perform PRIME process
Fill a sample tube with 1.2 ml of your sample. Check the tube for eventual imperfection or
contaminants (cracks, aggregates, hair) and replace or remove, resp. if necessary.
Insert the tube into the sample port, push it up until a distinct click is heard
Press the start button. In the console window, information on the status of the
measurement is displayed. The following work is proceeding step by step.
CyView
™StartPrerun
Prerun and stabilization of sample flow
Start Measure
Data acquisition
Console displaying control board status
During the data acquisition the total particle count and the analysis rate is indicated on the
real-time count meters