![Sylvac D 100S Instructions For Use Manual Download Page 67](http://html.mh-extra.com/html/sylvac/d-100s/d-100s_instructions-for-use-manual_1417043067.webp)
67
3 D104 4-CHANNEL RS232 SWITCHING UNIT
3.1 GENERAL DESCRIPTION
The D104 switching unit is designed to link up to four OptoRS simplex or duplex instruments to be connected to unit
D100S
, or, by connecting several
D104
units in parallel, up to 16 instruments. The
D100S
unit automatically reads
the values measured by the Sylvac RS 232 instrument and processes them in the same way as a value measured
by a probe.
The
D104
unit may be fixed to the
D100S
by inserting four plastic legs. It may also be installed separately, close to
OptoRS instruments (maximum RS232 connection distance is 15 meters).
3.2 CONNECTION OF 2 D104 UNITS AND 1 D108 UNIT ON D100S UNIT
3.3 OPERATION
The OptoRS simplex or duplex instruments are connected to inputs 1 to 4 of the D104 unit and correspond to
channels 1 to 4 on the display.
If a second
D104
unit is used, the inputs marked 1 to 4 on this unit correspond to channels 5 to 8 on the display.
This second unit will be connected in the same way as above : the ‘Extension’ input/ouput of the first
D104
unit is
connected to the ‘RS232’ input/ouput of the second
D104
by means of the 9p. male cable supplied. It is also
necessary to connect the output of the first
D104
to the ‘In’ input od the second
D104
.
Configure the OptoRS channels by pressing Setup then channel, then type in the OptoRS channel number followed
by Enter, then Enter if the channel is allocated to the OptoRS instrument connected to this channel number, then 2
for OptoRS simplex instruments or 3 for duplex.
It is now possible on the
D100S
display unit to enter presets, tolerances, choose the direction of measurement and
the radius/diameter mode for each channel. Once this has been done, it is possible to select manually the channel to
be displayed, and to change channels by means of the external contact (foot pedal) if this has been configured to do
so. These functions are explained in detail in Section 1.6.
Summary of Contents for D 100S
Page 2: ...2 ...
Page 76: ...76 Edition 2005 12 V1 4 SYL D100S E 681 038 110 ...