35
13/ Enter nominal size and tolerances for all channels to be scanned (refer to section 1.6.7).
14/ Put a block gauge (master) under the probes and be ready to calibrate the channels.
15/ Enter the preset values according to the master for all required channels. For a new calibration in scanning
mode, just press
Preset
when the master is in place.
16/ The system is now ready. Position the work piece to be measured and when the probes are in the measuring
position, press the
Scan
key.
According to the particular channel allocation specified in point 10, scanning will stop at channel 4 and the following
will be displayed :
Take a minimum measurement on channel 4 and then press
Enter
key or
foot pedal.
In case of a wrong measurement, the
Clear
key will allow a new
measurement to be taken.
17/ When scanning is completed, the following will be displayed (numerical value are given as examples) :
Sequential work-pieces numbered : from 1 to max 65 000.
General indication for the measured work piece from indicators lights.
This criteria takes into account the internal or external measurement for
each channel. (
refer to tolerances in section 1.6.7
)
Display of channel number, measuring mode if channel is in A +/- B mode,
And measured value or, if tolerances are active, of the difference between
nominal size and measured value and graphical display <, = or >.
The following keys are active in Scan mode :
Starts a new scan. Work piece numbering is done automatically (from 1 to n pieces). For 40 or 80
Columns printing format, it is also possible to enter an alpha-numerical identification of each scanned
une work piece (
refer to section. 1.6.14
:
key).
Displays or removes the tolerance indicators for each channel.
Tol not activated
: measured value is displayed.
Tol activated
: difference between nominal and measured value is displayed, followed by
E
for an
external measurement or
I
for an internal measurement, and finally by tolerance indicators
<
,
=
or
>
.
The first displayed values show a maximum of 5 channels. Any following channels will be shown by
Pressing this key. When the last table is reached ( the arrow disappears from the upper-right-hand
Side of the display), you can exit scanning mode by pressing this key again.
This key allows display of the preceding table. When the first table is reached ( the arrow
disappears) pressing the key again allows you to exit scanning mode.