1.6.14.5 Format for 15 columns printer
For small printers, battery or accumulator powered (e.g. EDP5000 or SP1).
As previously, the unit requests the time needed to print one line. A header cannot be entered here.
a/ in normal mode, the measurements are printed as in computer format, but are preceded by the channel number.
In tolerance mode, the difference between the nominal dimension ant the measurement is printed :
1 9.716
2+3 15.434
b/ in scanning mode, the following is printed ( in tolerance mode, the difference between the nominal dimension and
the measurement is printed) :
WORKPIECE 4 <
1 9.716
2 12.54
3 2.008
c/ printing of statistical values : as per computer format but with the possible addition of stored values :
d/ printing of stored values : they are preceded by the store number (figure from 1 to N ) :
1 9.999
2 9.998
3 10.002
4 10.000
1.6.14.6 Remote command of D100S unit :
Practically all functions of the D100S unit are remote controllable from a computer through input RS232, according
to the following general method :
The first 3 letters of functions are used for remote command. For example, if a measured value is required from the
computer, the first 3 letters of Print are transmitted, i.e.
PRI
.
Any number of spaces can be inserted anywhere, except inside numbers.
Command characters may be in upper or lower case (the latter are ASCII coded).
On/Off functions like analog scale or tolerances indicators are activated with the first 3 letters of the function,
followed by ON or 1. They are disabled with OFF or 0 (= zero and not the letter O).
For example : TOL ON displays tolerances indicators, as does TOL 1.
There should be no delay between characters in remote command word.
Words used for remote commands :
CHA+
(CHANNEL)
CHA-
CHA3
CHA-2
CHA+1-2+
CHA*2
CHA*1
CHA*1.5
CHA? or CHA PRI
CHA2=1
CHA1=1RS
CHA1=1RSD
CHA1=1
= selects positive measuring direction of the indicated channel.
= selects negative measuring direction of the indicated channel.
= selects channel 3.
= selects channel 2, in negative measuring direction.
= selects channel 1 minus channel 2, both channels with positive measuring direction.
= establishes diameter function of the selected channel. (X*2)
= establishes radius function (normal) of the selected channel. (X*1)
= 1,5 multiplication factor.
= identifies the direction of measurement (+/-) followed by the active channel number.
= channel 2 assigned to probe 1.
= channel 1 assigned to RS232 instrument (simplex cable)
= channel 1 assigned to RS232 instruments (duplex cable)
= channel 1 assigned to analog instruments (P5,P10…)