53
2/ Same instruments, but with correction of 10 points (every 2.5 mm—> max. error 0.7 um :
Introduction of correction : probe out to probe in.
1/ Switch off unit D100S
2/ Switch on unit while pressing
Store
key for at least 5 seconds.
3/ If there is room, the date of old correction is displayed.
4/ Insert value of first reference in keyboard, normally 0. In theory, this is the probe travel stroke, approx. 0.8 mm
for a
P25
probe. Confirm by pressing
Enter
key.
5/ Then enter correction points. Gauge block may be used, whose exact value is know. With probe in opposition on
the pad, insert its exact value on keyboard and confirm by pressing
Enter
key.
6/ Continue in same way for all correction points, with 1 point min. and 25 points max.
7/ When the final correction point is reached e.g. 5, press the
Clear
key and this will exit correction mode (otherwise
you will move to the next point.
Correction is indicated by an E on the display (E for Extended accuracy)
This value cannot be deleted, even if you restart the unit (Reset)
If you wish to delete the effect of the multiple correction, briefly press
Store
key when switching the unit on.
To reconfirm this correction, repeat the same sequence as above.
Correction is the same for all channels of the D100S. Multiplexing units D102 and D108 for multiple probe use are
invalid.