52
The date of the last calibration is shown.
6/ Place probe on reference base (value 0). Press
enter
Enter
key.
7/ Insert 2 mm test block for the
P2
, 5 mm for a the
P5
, a 10 mm
for the
P10
, 25 mm for a
P25
or 50 mm for a
P50
.
8/Enter exact value of pad on keyboard : 2, 5, 10, 25 or 50 mm, then
press
Enter
key.
1.7 CALIBRATION OF THE UNIT
1.7.1 General calibration
The D100S units are calibrated at the factory. However if a re-calibration is required, proceed as follows :
1/ Fix a
P2, P5, P10, P25
or
P50
probe to a vertical support.
2/ Select a resolution of 0.0001 mm or 0.00001 in.
3/ With the probe in its fully extended contact-free position, reset display to zero (
Preset
key).
4/ Mechanically position probe under reference base so that readout indicates a measurement :
for
P2
between 0.2 and 0.3 mm
P5
between 0.7 and 0.8 mm
P10
between 0.4 and 0.5 mm
P25
between 0.8 and 0.9 mm
P50
between 1.0 and 1.2 mm
5/ Switch off unit, then switch on by pressing
channel
key for four seconds. The following display will appear :
The calibration value is automatically stored and can only be deleted by a new calibration (changing lithium module
Or accumulator does not effect the calibration).
1.7.2 Coupling probe to unit
The user can couple the probe-unit couple and correct from 2 to 25 points linearly.
Example of correction :
1/ Probe P25 with D100S without correction —> max error 1.4 um :