AMI Turbitrack
Troubleshooting
45
E014
Case Temp. low
– check case/environment temperature
– check programmed value
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1 & 2
E018
Turbi disconnected
– shut off power
– check wiring
E024
Input active
– See If Fault = Yes is programmed in
E026
IC LM75
– call service
E028
Signal output open
– check wiring on signal outputs 1 and 2
E030
EEProm Frontend
– call service
E031
Calibration Recout
– call service
E032
Wrong Frontend
– call service
E033
Power-on
– none, normal status
E034
Power-down
– none, normal status
E068
Replace absorber
(dew point inside the
turbidimeter housing is
3 °C or higher)
– call the <Maintenance>/<Replace
Absorber> menu item
– replace the humidity absorber, see
Replacing the Humidity Absorber, p. 38
Error
Description
Corrective action