A-96.250.231 / 130421
AMI Turbitrack
Program List and Explanations
51
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
Provides the list of active errors with their status (active, acknowl
-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Message List
Shows the error history: Error code, date and time of issue, and sta
-
tus (active, acknowledged, cleared).
64 errors are memorized. Then the oldest error is cleared to save the
newest error (circular buffer).
1.3 Maintenance List
Provides the list of necessary maintenance. Cleared maintenance
messages are moved to the Message list.
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
2.1.1
Designation
: of instrument: AMI Turbitrack
2.1.2
Version
: firmware version, e.g. V6.24-01/20
2.1.3
Version TURBI3:
, e.g. 1.01
2.1.4
Factory test:
date of instrument, mainboard and frontend QC test.
2.1.5
Operating time:
years, days, hours, minutes, seconds
2.2 Sensors
2.2.1
Turbidity:
Turbidity
: in FNUor NTU
Raw value
: Turbidity in counts.
Quotient
: Division of detector signal by reference signal.
Scale factor 1
: correction factor 1 determined during factory calibra
-
tion.
Scale factor 2
: correction factor 2 determined during factory calibra
-
tion.