A-96.250.231 / 130421
AMI Turbitrack
Troubleshooting
44
Error
Description
Corrective action
E001
Alarm high
– check process
E002
Alarm low
– check process
E005
Range
– turbidity out of range
– disappears if measuring chamber is
filled
E006
Absorber exhausted
(dew point inside the
turbidimeter housing is
5 °C or higher)
– call the <Maintenance>/<Replace
Absorber> menu item
– replace the humidity absorber, see
Replacing the Humidity Absorber, p. 38
– if the error appears again after a few
days, proceed as follows:
Make sure that a new, unused hu
-
midity absorber has been installed.
Call the <Replace Absorber> func
-
tion again and wait for another three
days. During this time, observe the
dew point and make sure that it de
-
creases.
if the dew point does not decrease,
call service.
E009
Sample Flow high
– check sample flow
E010
Sample Flow low
– check inlet pressure
– re-adjust the sample flow
– clean instrument
E013
Case Temp. high
– check case/environment temperature