A-96.250.231 / 130421
AMI Turbitrack
Product Description
11
On-line
Operation
The sample flows through the sample inlet [B] and the sample inlet
valve [C] into the measuring chamber [A], where the turbidity is mea
-
sured. From there it flows though the pressure regulator [F] with
which the sample flow can be regulated. Then the sample flows via
capillary and flow meter [H] into the drain funnel.
To keep the measuring chamber clean and free from deposits it can
be cleaned manually or automatically:
manually by turning the lever [D] located on the solenoid valve
support upwards.
automatically in programmable time intervals in which the sole
-
noid valve [E] is activated, see
.
The small pressure wave caused by opening and closing the valve
whirls up deposited particles and flushes them out of the measuring
chamber the next time the valve is opened.
A
B
C
D
E
Measuring chamber
Sample inlet
Sample inlet valve
Manual cleaning lever
Solenoid valve for cleaning
F
G
H
I
Pressure regulator valve
Drain funnel
Flow meter
Capillary
A
B
C
D
F
H
I
G
E