AMI INSPECTOR Conductivity
Default Values
A-96.250.611 / 120617
63
10. Default Values
Operation:
Sensors:
Filter Time Const.: ................................................................. 10 sec
Hold after Cal.:..................................................................... 300 sec
Relay Contacts
Alarm Relay .................................................. same as in Installation
Relay 1and 2 ................................................ same as in Installation
Input.............................................................. same as in Installation
Logger:
Logger Interval:.............................................................. 30 Minutes
Clear Logger:................................................................................ no
Installation:
Sensors
Flow: ........................................................................................ None
Sensor Parameters; Cell Constant ...............................0.0415 cm
-1
Sensor Parameters; Temp. corr............................................ 0.00 °C
Sensor Parameters; Cable length ........................................... 0.0 m
Sensor Parameters; Meas. unit ............................................. µS/cm
Temp. Compensation; Comp. ................................................... none
Quality Assurance; Level ......................................................... 0: Off
Signal Output
Parameter:.................................................................... Conductivity
Current loop:................................................................ 0 / 4 –20 mA
Function:.................................................................................. linear
Scaling: Range low:............................................................0.000 µS
Scaling: Range high: ................................................................1 mS
Alarm Relay:
Alarm Conductivity:
Alarm high: ...........................................................................300 mS
Alarm low:...........................................................................0.000 µS
Hysteresis:............................................................................1.00 µS
Delay: ......................................................................................5 Sec
Sample Flow:
Flow Alarm.................................................................................. yes
Alarm high: ..............................................................................20 l/h
Alarm low:..................................................................................5 l/h
Sample Temp:
Alarm High:............................................................................ 160 °C
Alarm Low:................................................................................. 0 °C
Case Temp. high:..................................................................... 65 °C
Case Temp. low: ........................................................................ 0 °C