AMI INSPECTOR Conductivity
Program List and Explanations
A-96.250.611 / 120617
51
5.2 Signal Outputs
5.2.1
Signal Output 3 (signal outputs 1 and 2 are deactivated)
5.2.1.1
Parameter:
Assign one of the process values to the signal output.
Available values:
Conductivity
Temperature
Sample flow
Cond. uc
5.2.1.2
Current Loop:
Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
5.2.1.3
Function:
Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See
As process values, p. 51
Control upwards or control downwards for controllers.
See
As control output, p. 53
As process
values
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below.
A
B
linear
bilinear
X
Measured value
20
0.0
0.1
0.2
0.3
0.4
0.5
10 12
(0 - 20 [mA])
0 / 4
(4 - 20 [mA])
[mA]
X
A
B