AMI INSPECTOR Conductivity
Installation
A-96.250.611 / 120617
21
3.5.3
Relay Contacts 1 and 2
NOTICE:
For resistive loads only; do not use with capacitive or
inductive loads. Rated load 100 mA / 50 VAC.
For programming see Menu Installation
5.3.2 and 5.3.3, p. 62
3.6.
Signal Output
The optional signal output 0/4–20 mA PCB can be plugged onto
the USB interface PCB.
NOTICE:
Max. burden 510
Terminals 16 (+) and 15 (-).
For programming see menu
5.2 Signal Outputs, p. 55
Terminals
Description
Relay connection
NO
Normally
Open
6/7: Relay 1
8/9: Relay 2
Inactive (opened) during normal
operation and loss of power.
Active (closed) when a pro-
grammed function is executed.
A
B
Signal output 0/4–20 mA PCB
USB interface
A
B
6/8
0V
7/9