![Swan Analytical Instruments AMI INSPECTOR Conductivity Operator'S Manual Download Page 41](http://html1.mh-extra.com/html/swan-analytical-instruments/ami-inspector-conductivity/ami-inspector-conductivity_operators-manual_1411377041.webp)
AMI INSPECTOR Conductivity
Error List
A-96.250.611 / 120617
39
Error
Description
Corrective action
E001
Alarm high
– check process
– check programmed value, see
5.3.1.1.1,
p. 55
E002
Alarm low
– check process
– check programmed value, see
5.3.1.1.25, p. 55
E007
Sample Temp. high
– check process
– check programmed value, see
5.3.1.3.1,
p. 56
E008
Sample Temp. low
– check process
– check programmed value, see
5.3.1.3.25, p. 56
E009
Sample Flow high
– check Inlet pressure
– readjust sample flow
– check programmed value, see
5.3.1.2.2,
p. 56
E010
Sample Flow low
– check Inlet pressure
– readjust sample flow
– clean flow cell
– check programmed value, see
5.3.1.2.35, p. 56
E011
Temp. shorted
– Check wiring of sensor, see
Connection
Diagram, p. 17
– Check sensor
E012
Temp. disconnected
– Check wiring of sensor, see
Connection
Diagram, p. 17
– Check sensor
E013
Case Temp. high
– check case/environment temperature
– check programmed value, see
5.3.1.4,
p. 56