12
A-96.250.611 / 120617
AMI INSPECTOR Conductivity
Product Description
2.3.
Technical Data
Power Supply
Battery
Use original, supplied power adapter only.
Voltage:
Power consumption:
Charging time:
Battery type:
85–265 VAC, 50/60 Hz
max. 20 VA
6h
Li-Ion
During charging protect from heat impact and keep splash-proof
(not IP66).
Operating time
Stand-alone (Battery):
Connected adapter:
> 24h
continuous
Controlled shut-down when battery is empty, remaining time is dis-
played.
Electronics
Aluminium with a protection degree of IP 66 / NEMA 4X
housing
Ambient temperature:
Humidity:
Display:
-10 to +50 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
5–20 l/h
up to 50 °C
up to 2 bar
pressure free
On-site
The analyzer site must permit connections to:
requirements
Sample inlet:
Sample outlet:
Max. Altitude:
1/4” Swagelok tube adapter
flexible tube 6x8 mm
2000 m above sea level
Measuring
Range
Range:
0.055 to 0.999
s/cm
1.00 to 9.99
s/cm
10.0 to 99.9
s/cm
100 to 1000
s/cm
Resolution
0.001
s/cm
0.01
s/cm
0.1
s/cm
1
s/cm
Accuracy
Automatic range switching.
±1% of measured value ±1 digit
Values for the cell constant 0.0415 cm
-1
Swansensor
UP-Con1000 SL.