AMI INSPECTOR Conductivity
Program List and Explanations
A-96.250.611 / 120617
47
2.2.1.5
Cal. History:
Review diagnostic values of the last calibrations.
Only for diagnostic purpose.
Number
Date, Time
Cell constant
Max. 64 data records are memorized. One process calibration cor-
responds to one data record.
2.2.2
Miscellaneous
:
2.2.2.1
Case Temp:
Shows the current temperature in °C inside the trans-
mitter.
2.3 Sample
2.3.1
Sample ID
:
Shows the identification assigned to a sample. This identification
is defined by the user to identify the location of the sample.
Temperature
: Shows the current sample temperature in °C.
(Pt 1000)
: Shows the current temperature in Ohm.
Sample Flow
: Shows the current sample flow in l/h
(Raw Value)
in Hz.
The Sample flow must be above 5 l/h.
2.4 I/O State
Shows current status of all in- and outputs.
2.4.1/2.4.2
2.5 Interface
2.5.1
Protocol USB Stick.
Alarm Relay:
Active or inactive.
Relay 1 and 2:
Active or inactive.
Input:
Open or closed.
Signal Output 3:
Actual current in mA (if option is installed)