AMI INSPECTOR Conductivity
Program List and Explanations
A-96.250.611 / 120617
57
5.3.2 and 5.3.3
Relay 1 and 2:
The function of relay contacts 1 or 2 is defined by
the user.
NOTICE:
The navigation in the menu <Relay 1> and <Relay 2>
is identical. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1
First select the functions as:
-
Limit upper/lower,
-
Control upwards/downwards,
- Timer
- Fieldbus
2
Then enter the necessary data depending on the selected func-
tion. The same values may also be entered in menu
4.2 Relay
Contacts, p. 49
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20
Parameter:
select a process value
5.3.2.300
Setpoint
: If the measured value rises above respectively falls below
the set-point, the relay is activated.
5.3.2.400
Hysteresis:
within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
5.3.2.1
Function = Limit upper/lower:
Parameter
Range
Conductivity
0
S–300 mS
Temperature
-25 to +270 °C
Sample flow
0–50 l/h
Cond. uc
0
S–300 mS
Parameter
Range
Conductivity
0
S–300 mS
Temperature
0 to +100 °C
Sample flow
0–50 l/h
Cond. uc
0
S–300 mS