10
A-96.250.611 / 120617
AMI INSPECTOR Conductivity
Product Description
Fluidics
The flow cell QV-Flow UP-Con SL consists of the flow cell block [B],
the flow sensor [C] and the flow regulating valve [E].
The Swansensor UP-Con1000 SL [A] with integrated temperature
sensor is inserted into the flow cell block [B].
The sample enters at the sample inlet [F]. It flows through the flow
regulating valve [E], where the flow rate can be adjusted. Then the
sample flows through the flow cell block were the conductivity and
temperature of the sample is measured.
The sample leaves the flow cell block via flow meter through the
sample outlet [D].
A
B
C
Swansensor Up-Con1000 SL
Flow cell block
Flow sensor
D
E
F
Sample outlet
Flow regulating valve
Sample inlet
A
B
C
D
E
F