40
A-96.250.611 / 120617
AMI INSPECTOR Conductivity
Error List
E014
Case Temp. low
– check case/environment temperature
– check programmed value,
see
5.3.1.5, p. 56
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
see
5.3.2 and 5.3.3, p. 57
E024
Input active
– See If Fault Yes is programmed in Menu
see
5.3.4, p. 60
E026
IC LM75
– call service
E030
EEProm Frontend
– call service
E031
Calibration Recout
– call service
E032
Wrong Frontend
– call service
E033
Power-on
– none, normal status
E034
Power-down
– none, normal status
Error
Description
Corrective action