G
Gu
uiid
da
an
ncce
e a
an
nd
d m
ma
an
nu
uffa
accttu
urre
err’’ss d
de
ecclla
arra
attiio
on
n -- e
elle
eccttrro
om
ma
ag
gn
ne
ettiicc iim
mm
mu
un
niittyy
Surge
IEC 61000-4-5
+1 kV line(s) to line(s)
+2 kV line(s) to earth
P
Po
ow
we
err--P
PR
RO
O: N/A
S
SM
MR
RT
T Charger (6500-201-
010):
+1 kV line(s) to line(s)
+2 kV line(s) to earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11
0% U
T
for 0.5 cycle at 0°,
45°, 90°, 135°, 180°, 225°,
270°, and 315°
0% U
T
for 1 cycle
70% U
T
(30% dip in UT) for
25 cycles
0% U
T
for 250 cycles
P
Po
ow
we
err--P
PR
RO
O: N/A
S
SM
MR
RT
T charger (6500-201-
010): 0% U
T
for 0.5 cycle at
0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% U
T
for 1 cycle
70% U
T
(30% dip in UT) for
25 cycles
0% U
T
for 250 cycles
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the S
SM
MR
RT
T charger requires
continued operation during
power main interruptions, it
is recommended that the
device be powered from an
uninterrupted power supply
or a battery.
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Electrical transient
conduction along supply
lines
ISO 7637-2
per ISO 7637-2
P
Po
ow
we
err--P
PR
RO
O: N/A
S
SM
MR
RT
T Charger (6500-201-
010): per ISO 7637-2
N/A
N
No
otte
e:: U
T
is the a.c. mains voltage before applications of the test level.
6550-309-005 Rev H.0
59
EN