![Stryker Power-PRO TL Cot Operation Manual Download Page 59](http://html.mh-extra.com/html/stryker/power-pro-tl-cot/power-pro-tl-cot_operation-manual_1385991059.webp)
E
Evve
erryy ssiixx m
mo
on
ntth
hss o
orr 1
12
2 h
ho
ou
urrss
Inspect these items every six months or 12 hours, whichever comes first.
IItte
em
m
IIn
nssp
pe
ecctt
Hydraulics
Hoses and fittings for damage or wear
Hydraulic velocity fuse - place a weight of approximately 50
lb (23 kg) on the cot, raise the cot, lift the cot with two
operators, pull the manual backup release handle, rapidly
set the cot down, and make sure that the cot does not drop
Electronic controls
Extend cot to raised position, measure and check load
height
Jog function operates
High speed retract works
Switches
No damage or wear to either switch
Both switches operate
Litter
No bent, broken, or damaged components
No damage or tears on cot grips
Siderails operate and latch
Foot rest operates
Mattress
No cracks or tears
Base
No bent, broken, or damaged components
Wheels
Free of debris
Head section
No bent, broken, or damaged components
Grip bar has no excessive damage or tears
Load wheels are secure and roll
Kickstand (optional)
Lubricate the kickstand spring and internal spring housing
(optional) using T
Trrii--F
Fllo
ow
w® lubrication.
E
Evve
erryy 1
12
2 m
mo
on
ntth
hss o
orr 2
24
4 h
ho
ou
urrss
Inspect these items every 12 months or 24 hours, whichever comes first.
IItte
em
m
IIn
nssp
pe
ecctt
Settings
Cot and fastener fit and function
Safety bar connects to the vehicle safety hook
Cylinder
Cylinder is adjusted - lock nut is tight and the cot stops
moving when it hits the dead stops
Manual backup release handle
Returns to the stowed position
Litter
All welds are intact, not cracked, or broken
Warning labels present, legible
6550-309-005 Rev H.0
55
EN