Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 30 of 61
PIP - Quik-Combo Defibrillator a Positive R-Wave Test
To perform A positive R-wave test:
1. Set the Impulse 7000DP to a
1mV, Normal
Sinus, 60 BPM
.
2. Set the device
ECG SIZE
to
1.0
.
3. Set the
LEAD
selection to
PADDLES
.
4. Print 10 seconds of paddles ECG recorded on
printer paper.
5. Turn the printer
OFF
.
6. Confirm the R-Wave polarity on the paper
matches image as shown in Figure 1.18.
Figure 1.18: A Positive R-Wave