LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 13 of 61
PIP - User Test and Date/Time Verification Tests
To perform the User Test and Date/Time Verification Tests:
1. Connect the Quik-Combo and Test Plug into the device.
2. Turn the device
ON
.
3. Press
OPTIONS
to access the Options menu.
4. Select
USER TEST
from the Option menu. When asked to Start user test.
Select
Yes
and confirm using the Speed Dial.
5. Confirm the printed results that the device passes the user test and that the
correct date and time values are displayed on the printout as shown in
Figure 1.4.
Note:
If the date and time are incorrect, reset using the
Options/Date/Time menu.
6. Turn the device
OFF
.
Figure 1.4: User Test and Date/Time Printout