LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 24 of 61
PIP - Analog ECG Output Test (Optional)
Note:
Perform this test if this feature is used by the customer. Otherwise, skip to the next test.
To perform Analog ECG Output Test:
1. Connect the device to the Impulse 7000DP and oscilloscope
as shown in Figure 1.14.
2. Turn the device
ON
.
3. Using the ECG cable supplied with the device input a 1-mV,
10-Hz sine wave from the Impulse 7000DP.
4. Set device
LEAD
selection to
LEAD II
.
Note:
The Analog ECG output is in real time at a nominal
1 V/mV and is not affected by the device ECG SIZE
setting.
5. Verify the amplitude of the signal displayed on the
oscilloscope is between 0.90Vp-p and 1.10Vp-p.
6. Disconnect the ECG output cable from the device and
oscilloscope.
7. Turn the device
OFF.
Figure 1.14: Analog ECG Output Test Setup