Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 48 of 61
Perform leakage current testing in accordance to the following electrical safety standards:
IEC (International Electro technical Commission) 62353.
WARNING
SHOCK HAZARD
Failure to properly perform these tests could result in a failure to detect excessive leakage current. Make sure you are familiar
with your test equipment and these test performance procedures.
Leakage
– Current flow induced by the application of high voltage to a material or object with high dielectric strength.
Normal Condition (N.C.)
– AC voltage is applied in either normal or reversed polarity (that is, measurements made with the POLARITY
switch in both NORMAL [NC] and REVERSED [RM] positions). The earth ground is intact during these measurements.
Single Fault Condition (S.F.C.)
– AC voltage is applied in either normal or reversed polarity (that is, measurements made with the
POLARITY switch in both NORMAL [NC] and REVERSED [RM] positions). The earth ground is NOT intact during these
measurements.
Safety Analyzer setup instructions are specific to the Fluke Biomedical ESA612.