LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 14 of 61
To perform SPO2 Test:
1. Turn the device
ON
.
2. Connect the SpO2 finger probe to the SpO2 connector as shown in Figure 1.5.
3. Verify the
SpO2 parameter region
appears on the display.
4. Place your index finger into the SpO2 finger probe. Allow several seconds for the probe to
find your pulse.
5. Confirm the SpO2 reading is between
50 % to 100 %
.
6. Remove finger from SpO2 finger probe. Message should display on screen:
SpO2 Check
Sensor with audible alarm
.
7. Unplug SpO2 finger probe from the device; while on current patient record, message should
display:
SpO2 No Sensor Detected
.
Note
: Alarms will clear when device is turned off and new patient record started the next
time the device is turned on again.
8. Turn the device
OFF
.
Figure 1.5: SPO2 Test