LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 55 of 61
PIP-Direct Applied Part Leakage Test-Therapy
WARNING
POSSIBLE EQUIPMENT DAMAGE
Do not defibrillate when connected to the safety analyzer!
1. Turn the device ON.
2. Press the LEFT/RIGHT arrows to select the LL and LA leads, and set the Safety Analyzer controls as follows:
Note:
The A.P. group should be set up to measure the combined leakage of the LL and LA leads.
3. Press the TEST button to measure the Direct Applied Part Leakage current.
4. Verify the device AC Mains LED is ON.
5. Verify the measured current is between 2
A and 2625
A (120 VAC and 240 VAC).