PIP - Quik-Combo Defibrillator ECG Characteristic Tests
Note:
Perform this test if Quik-Combo cable is used. Otherwise, skip to the next test.
The ECG characteristic tests consist of ECG gain and a positive R-wave test. These two tests are included here as a single procedure
and step numbers are continuous from one step to the next.
PIP - QUIK-COMBO Defibrillator ECG Gain Test
PIP - QUIK-COMBO Defibrillator a Positive R-wave Test
PIP - Quik-Combo Defibrillator ECG Gain Test
To perform ECG Gain Test:
1.
Set the Impulse 7000DP output to
ECG
, Performance,
1-mv, 10-Hz sine
wave
.
2.
Set the device
ECG SIZE
to
4.0
.
3.
Set the
LEAD
selection to
PADDLES
.
4.
Print 10 seconds of paddles ECG and confirm the printed signal amplitude
is 36 mm to 44 mm, peak-to-peak as shown in Figure 1.17.
Figure 1.17: ECG Gain at Paddles Lead