Power Switch P/N 0700-001-412
Note
: Ensure that connections to the power switch are connected in the manner
below as improper connection can cause damage to equipment as well as harm
to technician.
1. Unplug the connections at each of the four prongs on the back of the
power switch
(P/N 0700-001-412) (AC). The prongs on the top have two wires in each connector,
while the prongs on the bottom only have one wire in each. The diagram below shows a
rear view of how the wiring is connected with the
power switch
(P/N 0700-001-412)
installed in the chassis.
Note
: If the
power switch
(P/N 0700-001-412) (AC) is replaced, the proper
electrical safety tests must be performed before placing docking station back in
service. Follow procedure
6.5 Testing
on page 6-20 to make sure the proper
testing requirements are met.
4-23
Summary of Contents for 0702-002-620
Page 1: ......
Page 3: ......
Page 5: ......
Page 15: ......
Page 17: ......
Page 19: ......
Page 21: ......
Page 22: ...Chapter 1 Operational Description...
Page 23: ......
Page 27: ......
Page 28: ...Chapter 2 Ultra Rover Disassembly...
Page 29: ......
Page 31: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 2...
Page 37: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 8...
Page 42: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 13...
Page 49: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 20...
Page 81: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 52...
Page 103: ......
Page 105: ......
Page 106: ...Chapter 4 Docker Disassembly...
Page 107: ......
Page 133: ......
Page 134: ...Chapter 5 Troubleshooting...
Page 135: ......
Page 167: ......
Page 168: ...6 33 Chapter 6 Technical Procedures...
Page 169: ......
Page 219: ......
Page 220: ...Chapter 7 Installation...
Page 221: ......
Page 224: ...7 3...
Page 228: ...Chapter 8 Appendices...
Page 229: ......
Page 231: ......