Earth Leakage Current Test (Covers Removed)
1. Plug the rover into the safety analyzer.
2. Turn on both the
Smoke Evacuator
and
Vacuum Control (for the small and large
canister)
and set both to maximum. (Canister suction and smoke evacuator to 100%)
NOTE:
To ensures that the measurement is taken with all functions at maximum, the
technician will need to measure and record the values in step 5 - while performing
step 4.
NOTE:
it is normal for the rover to turn off momentarily when the neutral switch is in
the open position. The rover functions may turn off with changes between each
setting below. Verify that required functions are running before recording the maximum
value for each combination requiring the neutral switch being placed in the closed
position.
3. Turn the analyzer dial to the LEAKAGE CURRENT position. The LEAKAGE switch
must be depressed toward the EARTH label to record the values in step 5.
4. Extend the IV pole by pressing the IV pole up button. (Hold through step 5)
5. Measure and record the earth leakage in all combinations listed in table 6.5.1.
Analyzer Settings
Leakage Current Limit
Polarity Neutral
Normal
Closed
20 uA < X < 300 uA
Reverse
Closed
20 uA < X < 300 uA
Normal
Open
20 uA < X < 1000 uA
Reverse
Open
20 uA < X < 1000 uA
Table 6.5.1 Analyzer Settings and Earth Leakage Current Limit for Rover
Patient Leakage (Covers Removed)
1. Plug the rover into the safety analyzer.
2. Attach the test lead to the fluid level transducer just above the plastic where it attaches
to the small canister cap.
3. Turn on both the
Smoke Evacuator
and
Vacuum Control (for the small and large
canister)
and set both to maximum. (Canister suction and smoke evacuator to 100%)
NOTE:
To ensure that the measurement is taken with all functions at maximum, the
technician will need to measure and record the values in step 6 - while performing
step 5.
4. Configure the safety analyzer to measure PATIENT LEAKAGE.
5. Extend the IV pole by pressing the IV pole up button. (Hold through step 6)
6-20
Summary of Contents for 0702-002-620
Page 1: ......
Page 3: ......
Page 5: ......
Page 15: ......
Page 17: ......
Page 19: ......
Page 21: ......
Page 22: ...Chapter 1 Operational Description...
Page 23: ......
Page 27: ......
Page 28: ...Chapter 2 Ultra Rover Disassembly...
Page 29: ......
Page 31: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 2...
Page 37: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 8...
Page 42: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 13...
Page 49: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 20...
Page 81: ...4 Remove front cover assembly P N 0702 001 070 CH and place aside 2 52...
Page 103: ......
Page 105: ......
Page 106: ...Chapter 4 Docker Disassembly...
Page 107: ......
Page 133: ......
Page 134: ...Chapter 5 Troubleshooting...
Page 135: ......
Page 167: ......
Page 168: ...6 33 Chapter 6 Technical Procedures...
Page 169: ......
Page 219: ......
Page 220: ...Chapter 7 Installation...
Page 221: ......
Page 224: ...7 3...
Page 228: ...Chapter 8 Appendices...
Page 229: ......
Page 231: ......