Error message
Cause
Solution
1 = E01: NO VRAM ERROR
Parameters not stored successfully
Switch device off/on. Default values are loaded.
When not successful, send device for repair
2 = E02: MOTOR OUTPUT FAILURE
Overload on the motor controller
Switch device off/on. If not successful, send
device for repair
3 = E03: –
4 = E04: –
5 = E05: –
6 = E06: –
7 = E07: –
8 = E08: –
9 = E09: –
10 = E10: –
11 = E11: NO FOOTSWITCH
Footswitch not connected correctly.
Check footswitch connection.
12 = E12: NO MOTOR IN
M1
Motor/handpiece not connected correctly.
Connect motor/handpiece correctly.
13 = E13: TOO HIGH TEMPERATURE MOTOR
M1
Motor/handpiece overheating.
Allow connected handpiece to cool
or operate another handpiece on 2nd channel.
Note:
A new handpiece cannot be used
immediately on the same channel!
14 = E14: NO MOTOR IN
M2
see E12.
see E12.
15 = E15: TOO HIGH TEMPERATURE MOTOR
M2
see E13.
see E13.
16 = E16: IRRIGATION PUMP NOT CLOSED
“Pump lever” not closed correctly.
Move lever to vertical position.
17 = E17: MOTOR BLOCKED OR WRONG
M1
Connected handpiece does not respond
Connect suitable handpiece
the selected mode
or
or
handpiece is blocked for a short time
release foot pedal, take load off handpiece and
when starting
start again.
18 = E18: MOTOR BLOCKED OR WRONG
M2
see E17.
see E17.
19 = E19: MOTOR STOPPED
M1
Handpiece blocked during operation.
Release foot pedal, take load off handpiece and
start again.
20 = E20: MOTOR STOPPED
M2
see E19.
see E19.
21 = E21: NO SCB MODULE
Send device for repair
Error table UNIDRIVE
®
Technische Beschreibung
Fehlersuche
Technical description
Troubleshooting
Техническое описание
Выявление неисправностей
Summary of Contents for 20711620-1
Page 2: ......
Page 5: ...III Ger teabbildungen Photographs of the equipment...
Page 6: ...IV...
Page 83: ...75 Blockschaltbild Block diagram Technische Beschreibung Technical description SCB...
Page 102: ......
Page 105: ......
Page 107: ......