Prism6 & Crystal Reader software User Manual
7
Grounding point
This point is connected to the earth. Use this connection if grounding is
required.
Caution
– attention required
o
The Prism6 instrument must be placed on a stable surface that is clean and
vibration-free.
o
Do not block the area in front of the Prism6 instrument and ensure that the
movable tray has sufficient space to move while the Prism6 instrument flap is
opening and closing.
o
A sufficient clearance area around the Prism6 instrument is required to
allow access to the main power switch.
o
A 10cm (4 inches) clearance must be left at the back of the device to ensure
proper ventilation.
o
Before transportation, the respective tray holder (Sapphire chips & Opal
chips) must be removed.
•
Additional warning for the Prism6 instrument
Warning- heavy object
Do not try to lift the Prism6 instrument alone.
Two persons are required to lift the Prism6 instrument.
•
General safety instructions for the Prism6 instrument:
o
Please ensure that the Prism6 instrument is operated as instructed in the provided User
Manuals. User Manuals are subjected to changes. The latest version can be accessed at the
Technical Resources webpage.
o
Stilla Technologies cannot be held responsible for any damages or injuries arising from
improper operation.
o
Do not use the Prism6 instrument if any parts are broken, chipped, rusty, or if the power cables
are damaged in any way.
o
Do not open the housing of the Prism6 instrument if not explicitly instructed by a Stilla
Technologies representative.
o
Opening the Prism6 instrument housing may breach any warranty for the Prism6 instrument.
o
Only operate the Prism6 instrument with the provided and specified detachable main supply
cords. Do not replace the components with non-specified cords.
o
Do not attempt any repairs or alterations except as expressly described in this User Manual or
as instructed by a Technical Support Representative.
o
Always disconnect the instruments from their power source before cleaning or moving the
Prism6 instrument.
o
For the Prism6 instrument, the mains supply source must meet the national regulatory
requirements.
o
The power cord of the Prism6 instrument must be connected to a wall outlet with a grounded
conductor.
o
For the Prism6 instrument, the mains voltage must correspond to the range given in the product
specification.