6-18
P129394-020
Operator Manual
Sterilizer Operation
Table 6-3. Factory-Set Optional Liquid Cycle and Cycle Values
The AMSCO 400 Series Sterilizer is equipped with the following optional factory programmed liquid sterilization
cycle and cycle values.
Cycles
Sterilize Temp.
Sterilize
Time
Dry
Time
Recommended Load
(by Chamber Size)
Validation
Standard
Liquid
*
250°F
(121°C)
45 Min.
N/A
16 x 16"
– 15 x 1,000 mL Flasks
20 x 20"
with Loading Car –
32
x
1,000 mL Flasks
with Rack and Shelves –
28
x
1,000 mL Flasks
ST-8: 2008
*
The liquid cycle is for non-patient contact use only.
Figure 6-9. Liquid Cycle In-Cycle Screens
Cycle Selection Screen
Cycle Selection Screen
Condition Phase
Sterilize Phase
Exhaust Phase
NOTE: Countdown time on display is estimated; timer self-corrects estimated time at beginning
of each phase.
Press for More Cycles
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...