9-27
Troubleshooting
Operator Manual
P129394-020
Alarm
Description
Screen with Operator Instructions
9.5.7 Door Seal A (or
B) Switch Malfunction
Occurs if door seal
switch A (or B) is closed
out of cycle.
Screen with Service Instructions
NOTE:
Pressure switch 2 (PS2)
will be shown for door B
seal.
STATUS . . ALARM!
286
DOOR A SEAL SWITCH MALFUNCTION
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
287
DOOR A SEAL SWITCH MALFUNCTION
-> SEAL SWITCH CLOSED OUT OF CYCLE
CAUSES AND CORRECTIONS:
1. SEAL SWITCH MALFUNCTION
-> CHECK PS1 FOR FAILED CLOSED.
2. SHORTED CABLE
-> CHECK CABLE AT P19 FOR SHORT CIRCUIT
PAPER
FEED
EXIT
CLEAR
ALARM
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...