9-25
Troubleshooting
Operator Manual
P129394-020
Alarm
Description
Screen with Operator Instructions
9.5.5 Recorder
Temperature Probe
Failure
Occurs if output from
recorder temperature
probe is outside normal
range.
Screen with Service Instructions
STATUS . . ALARM! RECORDER
271
TEMPERATURE PROBE FAILURE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
272
RECORDER TEMPERATURE PROBE FAILURE
-> RTD PROBE, RTD4, OUTPUT IS OUTSIDE
NORMAL RANGE
CAUSES AND CORRECTIONS:
1. LOOSE CONNECTION IN PROBE WIRING
-> REPAIR
2. PROBE FAILED
-> REPAIR
-> RECALIBRATE
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4. MAIN CONTROL FAILURE
-> CHECK CONTROL BOARD POWER STATUS LEDs
-> REPLACE CONTROL BOARD
-> RECALIBRATE
PAPER
FEED
EXIT
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...