7-27
Cycle and Control Value Programming
Operator Manual
P129394-020
7.9.2 Set up Generator Flush
when Automatic Utilities
Control Enabled
Auto utility control must be highlighted on utility control screen (#39).
Important:
Sterilizer must be in auto utility shutdown screen (#40)
for flush to occur. Flush cycle will not occur if sterilizer is at
STANDBY
screen (#0) at scheduled time.
Select
GENERATOR FLUSH CONTROL
at screen #49.
•
Control advances to automatic generator flush cycle
scheduling screen (#49) when Automatic Utilities Control is
enabled.
•
Screen #49 allows user to set up duration time. Default
duration time is 10 minutes.
1.
Generator flush time is always automatic when
AUTO UTILITY
SHUTDOWN
is enabled. Flush automatically begins one hour
before restart time scheduled in auto utility shutdown. See
for
information on setting restart time of Automatic Utilities Control.
2.
Use keypad to enter
FLUSH
duration time (5, 10 or 15 minutes).
Press
EXIT
to return to
UTILITIES CONTROL
screen (#39).
NOTE: If less than one hour before restart time when unit powers
up, then a flush cycle will not occur on that day.
Figure 7-10. Generator Flush Duration Time Control Screen
when Auto Utilities Enabled
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...