9-8
P129394-020
Operator Manual
Troubleshooting
Alarm
Description
Screen with Service Instructions
9.2.9 Too Long In
Exhaust
Occurs if chamber does
not exhaust to 4 psig
within the allotted time.
Screen with Service Instructions
Alarm
Description
Screen with Operator Instructions
9.2.10 Exhaust Rate
Too Fast
Occurs if liquid cycle
slow exhaust rate is too
fast.
STATUS . . ALARM!
200
TOO LONG IN EXHAUST
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPETE CYCLE
•
EXTEND EXHAUST TIME
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
201
TOO LONG IN EXHAUST
-> CHAMBER DID NOT EXHAUST TO ATMOSPHERIC
PRESSURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. CHAMBER DRAIN STRAINER PLUGGED
-> CLEAN
2. SOLENOID VALVE MALFUNCTION
-> REPAIR S03
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
FEED
SERVICE
HELP
STATUS . . ALARM!
241
EXHAUST RATE TOO FAST
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
TRY TO EXHAUST CHAMBER ACCORDING TO OPTIMAL
COOLING RATE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...