AC Mains Voltage Selection
AC power input voltage configuration (115 or 230 VAC) for the
receiver housing is determined by the position of a selector switch in
the fuse assembly. The currently selected voltage configuration is
identified in a small window on the fuse assembly.
To change the voltage selection
1
Disconnect the AC power cord.
Pry open the fuse compartment located directly below the AC
power switch on the back of the receiver housing and, with a
small screwdriver, pry out the fuse assembly.
Figure 2-8 Changing the voltage selection
2
Remove the voltage selection strip, rotate it 180° and insert it
back in with the correct AC line voltage showing (115 or 230) in
the fuse assembly window.
Figure 2-9 Voltage selector re-inserted
Note:
After changing the voltage selection, ensure that the correct
fuses are installed. Refer to
on page 2-9 for
the required fuse type.