Lead Off Comparators
•
These determine if lead fault has occurred.
•
They provide inputs to the ALS PAL to indicate lead fault.
Differential Amplifiers
•
These provide common mode rejection.
•
They eliminate ECG electrode offsets (up to ±300 mV).
Pacer Blanking Switches
•
These switches strip the detected pacer pulse.
Bessel Filter
•
This filter defines the system’s 30-Hz bandwidth.
•
It provides a time delay to facilitate pacer blanking.
•
It prevents ringing due to transient inputs.
Final ECG Amplifiers
•
These amplifiers provide most of the ECG signal gain.
DC Restoration Comparators
•
These comparators become active when the ECG output is
outside the dynamic range of the pulse-width modulator (PWM)
converter.
•
They define the 0.05-Hz high pass pole frequency.
•
They eliminate the ECG electrode offsets (up to ±300 mV).
Pacer Detection and Pacer Flag Generator
•
The pacer detector triggers the pacer flag generator whenever a
pacer signal is detected.
•
The pacer flag is transmitted to the receiver and controls the
pacer blanking switches.
Automatic Lead Switching PAL
•
The automatic lead switching PAL decides which lead should be
driven and closes the appropriate lead drive switch.
•
It provides logic to the pacer detector when selecting the ECG
vector.