Current Drain Test for 91341/91343/91347 Transmitters
1
Refer to
on page 4-40 for an illustration of the test
setup for the 91341/91343/91347 transmitter.
Note:
Be sure to install the ECG simulator to the 91341. If the ECG
simulator is not connected, the lead fault LEDs will cause the
display to pulse, and the 91341 will fail the test.
2
Verify that the current drain is less than 8 mA for the 91341 or
91347, or less than 13.5 mA for the 91343.
Figure 4-23 Test setup for the 91341/91343/91347
Any transmitters that fail the current drain test should be retested
using the complete current drain test procedure (refer to
on page 4-27). Defective transmitters
must be returned to ESC (factory repair) for servicing.