The current setting of the SpO
2
averaging period can be displayed
by pressing the ECG CHANNEL FORMAT key and enabling SpO
2
.
Figure 1-3 SIP switch setting in battery compartment
The sampling interval selection enables you to determine how often
an SpO
2
measurement will be taken. Less frequent SpO
2
readings
can extend the usable life of the battery. (Refer to
on page 1-12 for more information.) This
selection is made by setting SIP switches 3 and 4 beneath the
battery compartment. The default setting is CONTINUOUS.
1 2 3 4
SpO
2
averaging
period
SpO
2
reading
interval
5 6 7 8
ON
OFF
enable
normal
enable
disable
IABP
disable
NIBP
Adult
enable
NIBP
enable
IABP
enable
Neonate
enable
Service use
operation
Caution:
No SpO
2
monitoring occurs between episodic sampling intervals.
Clinical practice or medical judgement should be used in selecting
continuous or episodic SpO
2
monitoring mode for each specific
patient.
Table 1-5 SIP switch 3 and 4 Settings
SIP switch 3
SIP switch 4
Effect
OFF
OFF
Continuous sampling (default)
OFF
ON
2-minute sampling interval
ON
OFF
5-minute sampling interval
ON
ON
30-minute sampling interval
Caution:
SIP switch 8 must remain OFF for normal operation.