4 - 3 9
U
L T R A V I E W
D
I G I T A L
T
E L E M E T R Y
- S
E R V I C E
M
A N U A L
M
A I N T E N A N C E
Current Drain Test for 90341 and 90347 Transmitters
1
Refer to
for an illustration of the correct test setup
for transmitter models 90341 and 90347.
2
Verify that the current drain is less than 8 mA for the 90341and
90347 transmitters.
Figure 4-21 Test setup for the 90341 and 90347
Current Drain Test for 90343 Transmitters
1
Refer to
for an illustration of the test setup for the
90343 transmitter.
Notes:
•
Be sure to install the SpO
2
cable and sensor.
•
NIBP is not used.
2
Allow the 90343 stabilize for approximately 10 seconds. The
multimeter measurement will initially rise to approximately 15 mA,
and then lower to less than 13.5 mA. Verify that the current drain
is less than 13.5 mA for the 90343 transmitters.
Figure 4-22 Test setup for the 90343