Using the Johanson 8762 tuning tool (or equivalent), slowly
adjust C10 to center the frequency symmetrically between the
two sideband peaks. This is a coarse adjustment. Refer to
5
Using a non-metallic tuning tool, gently tap C10. If the center
frequency changes or drifts, return transmitter to the Spacelabs
Healthcare ESC for repair or replacement of C10.
6
Using a tuning tool with a wider blade, slowly adjust R9 to center
the frequency between the two side band peaks. Adjustments are
3 kHz CF to peak on wide-band transmitters (90343 and 90347)
and 1.5 KHz on narrow-band transmitters (90341).
Figure 4-13 Large shield with adjustment labels